Function Description
Figure 20 Spontaneous Breathing (Inspiration) - Typical for Breathing System P/N 4116398 or 4117529
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
The patient inhales the gas mixture (expiratory gas and fresh gas) from the
breathing bag. The gas mixture flows through the fresh-gas decoupling valve,
the inspiratory valve, the O2 sensor, the inspiratory hose, and through the Y-
piece into the lung. The pressure sensor measures the airway pressure.