FABIUS TIRO
8.8.2.4
8.8.2.5
8.8.2.6
Passive Scavenger Functional Test
8.8.2.7
8.8.2.8
8.8.2.9
8.8.2.10 Set the COSY breathing system APL valve to SPONT.
8.8.2.11 Open the oxygen flow control valve to a flow of 10 L/min and
8.8.2.12 After the breathing bag inflates, the absorber systems' breathing
8.8.2.13 Systems that are not equipped with a pressure gauge, observe the
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
If necessary, remove the filter for cleaning. Brush any accumulated
lint of dust off the filter. The filter can be further cleaned with a low
flow of clean air or oxygen.
Place the filter back, making sure that it lays flat on the valve port
orifice.
Reinstall the anti-occlusion cage on the scavenger body, making
sure the filter is properly seated.
Check for moisture accumulation in the breathing and scavenger
hoses. Remove and moisture found.
Short-circuit the COSY breathing system inspiratory and expiratory
valves with a 2 mm breathing hose.
Install a breathing bag on the COSY breathing system.
occlude the passive scavenger exhaust port connection.
pressure gauge must indicate a pressure of less than 5 cm H2O.
generated pressure waveform, set the threshold to 5 cm H2O. The
pressure waveform shall not rise and remain above the dotted
threshold trace during this performance test.
25
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)