Dräger Medical Fabius Tiro Technical & Service Manual page 282

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

PMS PROCEDURE (continued)
8.11.3.2 Detach the hose from the Pressure Test adapter (P/N 4114807)
8.11.3.3 Attach the appropriate O2 adapter to the O2 connector on the inlet
TEST FITTING
4109318
( )
8.11.3.4 Open an O2 (reserve) cylinder valve. What is the flow as indicated
N2O Pipeline Check Valve (if applicable)
8.11.3.5 Attach the appropriate N2O adapter to the N2O connector on the
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
gauge assembly and attach it to the inlet of the test flowmeter as
shown.
block, and connect the other end of the hose with the quick
disconnect fitting to the adapter as shown.
on the test flowmeter? ___cc
inlet block, and transfer the end of the hose with the quick
disconnect fitting to the N2O adapter.
40
RETURN TO CD-ROM TABLE OF CONTENTS
QUICK DISCONNECT
5 cc/min.
FABIUS TIRO
S000081
TEST
FLOWMETER
O2 ADAPTER
SV88021