Fabius Tiro
Figure 6
Boot Strap Jumper and Download Connections on Control PCB P/N 4116632
232
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
13. Cycle machine power and verify all self-diagnostics indicate a PASS con-
dition. NOTE: NVRAM settings will be set when step
14. Reinstall the back panel of the control unit.
15. Access the appropriate service screens (see the
Diagnostics Section) and update the machine serial number, clear service
log and reset hours run. The machine serial number is located on the rear
of the machine.
16. Perform PMS test Steps Calibrations, Site Configurations, Oxygen Moni-
tor, Oxygen Concentrations, Pressure Monitor and Ventilator. Verify the
correct market kit settings are entered per original machine configuration
or customer requirements.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Software Update Procedure
General
Revision B Released
16
is performed.