PMS PROCEDURE (continued)
8.10.2.3 Are the cylinder gauges at zero (0) PSI? ___ (Y)
8.10.2.4 Open the cylinder valves.
8.10.2.5 Do the cylinder pressure gauges respond properly? ___ (Y)

8.11 Gas Inlet Regulator Output

8.11.1
O2 Inlet Regulator
8.11.1.1 Trolley or Wall Mount - Service Access
8.11.1.2 Bleed all cylinder and pipeline pressures. Disconnect all cylinders
8.11.1.3 Remove the pneumatic hose from the O2 outlet of the gas inlet
SV88016
TO N2O
FLOW CONTROL
VALVE
TO O2
FLOW CONTROL
VALVE
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
8.11.1.1.1 Fully extend the writing tray to its out position.
8.11.1.1.2 Remove the top drawer by releasing the locking tabs located
in the drawer slides (x2). (Trolley Mount Only)
and pipeline hoses.
block and interconnect a digital manometer as shown.
TEST 'T'
CONNECTOR
M32803
38
RETURN TO CD-ROM TABLE OF CONTENTS
N2O OUTLET (TO N20
FLOW CONTROL VALVE)
GAS INLET ASSEMBLY
FLOW CONTROL VALVE)
FABIUS TIRO
N2O INLET
REGULATOR
ADJUSTMENT
SCREW
O2 INLET
REGULATOR
O2 OUTLET (TO O2