Fabius Tiro
Figure 29 Sectional View of the Safety Valve
8.2
Negative Pressure
Relief Valve
Figure 30 Sectional View of the Negative Pressure Relief Valve
9
Pneumatic System
40
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
The negative pressure relief valve allows the patient to spontaneously
breathe ambient air should the medical gas supply and/or Fabius Tiro fail.
The opening pressure of this valve is -8 mbar.
The pneumatic assembly provides pressure for the PEEP valve control, and
also provides vacuum for the ventilator bag-type rolling seals and the APL
bypass valve control.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description
Revision B Released