Fabius Tiro
Figure 4
Sensitive Oxygen Ratio Controller (SORC) Adjustment
5
Oxygen Sensor Cali-
bration
214
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Adjustment and Calibration Procedures
1. Set the main power switch to ON.
2. Access the main service screen (ref.
3. Scroll to Calibration and select O2 to bring up the O2 Zero calibration
screen. See
Figure
5.
6002.002
RETURN TO CD-ROM TABLE OF CONTENTS
Main Service Screen
Revision B Released
Section).