Diagnostics
4
Preventive Mainte-
nance
4.1
Activate Preventive
Maintenance Date
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
When the Preventive Maintenance category is selected, the screen shown in
Figure 9
will appear.
Turn the rotary knob to select Activate, then press the knob to confirm. The
next preventive maintenance due date can then be set using the screen as
shown in
Figure
10.
Figure 9
Preventive Maintenance Category Selected
With Activate selected, the screen will appear as shown in
the rotary knob to position the box under Mth and press the knob to highlight
the box. Turn the knob to select the next month preventive maintenance is
due, and press the knob to enter the month.
With the box under Day, press the knob to highlight the box. Turn the knob to
select the desired day, then press the knob to enter the day.
With the box under Year, press the knob to highlight the box. Turn the knob to
select the year, and press the knob to enter the year.
6020.002
Fabius Tiro
Figure
Revision B Released
10. Turn
101