7.2.4.1 Ambient Pressure

7.2.4.2 Plateau-Mean Display
Screen
128
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
To return to the Configure category screen, scroll to the return arrow and
press the knob.
When Ambient Pressure is chosen, the screen will appear as shown in
36. Turn the rotary knob to select Adjust, and press the knob. Then turn the
knob to set the correct ambient pressure in mbar for the location the unit will
reside, and press the knob to enter the value.
Using the Site Configurations table found in the PMS section, set the ambient
pressure (mbar) in accordance with the local elevation. Refer to the Site Con-
figurations portion of the PMS section for additional information.
NOTE
It is imperative the ambient pressure setting is properly set. Otherwise,
erroneous Vt settings could result.
Changes in barometric pressure resulting from local weather conditions have
minimal to no affect on Vt readings. Therefore, it is not necessary to change
the ambient pressure setting with changing local weather conditions.
Figure 36 Ambient Pressure Screen
To return to the Pressure category screen, scroll to the return arrow and
press the knob.
The breathing pressure monitor window can display Mean airway pressure or
Plateau airway pressure.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Revision B Released