a) Open the cylinder valve and check for a pressure indication on the
corresponding cylinder pressure gauge.
– O2: 1000 psi
– N2O : 700 psi
b) Close the cylinder valve and remove the cylinder from the yoke.
c) For any gas, the pressure should not drop more than 50 psi in two
minutes.
10. Invert unit into its upright position by performing steps given in the
Module Inversion Reassembly
11. Perform the PMS Procedure.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement Procedure
Section.
Revision B Draft
Core