Dräger Medical Fabius Tiro Technical & Service Manual page 138

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Fabius Tiro
132
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Figure 40 Serial Port Parameters Screen
Table 11
Values for Parameters
Item
BAUD RATE
PARITY
STOP BITS
DATA BITS
PROTOCOL
NOTE
It is important to ensure that communication protocols selected on each
host and external device are correct. Vitalink and Medibus protocols are
similar and if not set identically on each device, inaccurate data may be dis-
played on the remote device.
To return to the main service screen, scroll to the return arrow and press the
knob.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Selections
1200, 2400, 4800,
9600, 19200, 38400
NONE, ODD, EVEN
1 or 2
7, 8
VITALINK or MEDIBUS VITALINK
Revision B Released
Diagnostics
Factory Default
9600
NONE
1
8