Replacement Procedure
23
Spirolog Sensor and
Cable
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1. Set machine power switch to OFF and disconnect AC power from the
machine.
2. Remove the expiratory hose terminal from the breathing system, and
slide the Spirolog sensor out of the breathing system.
3. Disconnect the cable from the sensor by squeezing the connector at the
points opposite the latch while pulling the connector away from the sen-
sor.
NOTE
If the Core Unit is wall mounted, refer to the
Procedure
prior to performing the following step.
4. Remove the back panel from the controller housing, and disconnect the
Spirolog cable from J18 on the controller PCB.
5. Loosen the nut on the strain relief bushing at the interface panel. Remove
the strain relief bushing from the panel while supporting the rear nut.
6. Install the replacement Spirolog strain relief bushing into the interface
panel, and tighten the nut.
7. Connect the replacement Spirolog cable to J18 on the controller PCB.
8. Route the cable from the machine to the breathing system in the same
manner as the original.
9. Connect the outboard end of the Spirolog cable to the flow sensor. Note
that the sensor and connector are keyed.
10. Reinstall the back panel on the controller housing, and restore AC power
to the machine.
11. Perform the PMS procedure.
6020.002
Fabius Tiro
Wall Mount Service Access
Revision B Released