Electromagnetic Testing and Compliance

The following CAUTIONS and WARNINGS are applicable to the Fabius Tiro
Anesthesia Machine:
CAUTION
Do not use Fabius Tiro in the environment of NMR tomography equip-
ment. Malfunctions may result, thereby endangering the patient.
CAUTION
The use of portable and mobile radio frequency communications
equipment can affect medical electrical equipment. Do not use mobile
phones within a distance of 10 meters from the machine. Mobile
phones can cause malfunctions in electrical medical equipment,
thereby endangering the patient and the operator.
WARNING
No third-party components shall be attached to the anesthesia
machine, ventilator, or breathing system (except for certain approved
exceptions). For more information, contact your local Authorized Ser-
vice Organization or DrägerService at:
DrägerService
Draeger Medical, Inc.
3122 Commerce Drive
Telford, PA 18969
Tel:(215) 721-5402
(800) 543-5047
Fax:(215) 721-5784
CAUTION
Although the Fabius Tiro is designed to minimize the effects of ambi-
ent radio-frequency interference, machine functions may be
adversely affected by the operation of electrosurgical equipment or
short wave or microwave diathermy equipment in the vicinity.
CAUTION
Communications with external equipment may be temporarily
affected by electromagnetic interference due to the use of electrosur-
gical equipment.
Tables 1 through 4 contained in this section are provided for informational
purposes and general guidance, as required by IEC 60601-1-2:2001, Medical
Electrical Equipment, Part 1-2, General Requirements for Safety - Collateral
Standard: Electromagnetic Compatibility.
The Fabius Tiro is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of the Fabius Tiro should assure that it
is used in such an environment.
6020.002
Revision B Released
53