Oxygen Supply Pressure Alarm Switch Adjustment - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Adjustment and Calibration Procedures
Figure 2
Setting Inlet Regulators
3
Oxygen Supply
Pressure Alarm
Switch Adjustment
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NOTE
If the Core Unit is wall mounted, refer to the
Procedure
prior to performing the following step.
1. Wall Mount Access (If Applicable)
a) Regulator access can be obtained from the underside when the unit
is rotated 180°.
2. Trolley Mount Access (If Applicable)
a) Remove the top drawer by releasing the locking tabs located in the
drawer slides (x2).
b) Fully extend the writing tray to its out position.
3. Invert unit by performing steps given in the
sembly
Section.
6002.002
Fabius Tiro
Wall Mount Service Access
Core Mount Inversion Disas-
Revision B Released
209