6.4
PEEP Valve Calibra-
tion
112
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Figure 20 O2 Zero Calibration Screen
Press the Exit key to return to the Calibration category screen, or the Standby
key to return to the Standby screen.
When PEEP is selected, the screen shown in
NOTE
Before performing the PEEP valve calibration, a valid pressure calibration
must be performed. Otherwise, an 'Inv pres.cal' message will appear.
Follow the on-screen instructions to perform the calibration. During calibra-
tion, an "In Progress" message is displayed. When the procedure is complete
a PASS or FAIL message is displayed. The Calibrate and Exit keys are
located just below their legends at the bottom of the screen.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Figure 21
Revision B Released
Diagnostics
appears.