Dräger Medical Fabius Tiro Technical & Service Manual page 198

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Fabius Tiro
Figure 24 Fresh Gas Flow Sensors and Filter Assembly
192
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
13. Place the flow meter opening squarely over the O-ring and twist the flow
meter side to side until it is seated onto the manifold. Do not force the flow
meter over the O-ring or it will not seat properly.
14. Reinstall the two top screws and two underside screws and carefully reat-
tach the Flow Sensor Housing Cover to the Flow Sensor Housing.
15. Reinstall the flow meter bezel assembly in the machine (refer to the
Meter Bezel Assembly
16. Reconnect the pipeline and cylinder supplies and restore power to the
machine.
17. Perform the PMS procedure including calibrations.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement Procedure
instructions).
Revision B Released
Flow