Dräger Medical Fabius Tiro Technical & Service Manual page 295

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

FABIUS TIRO
( )
8.19.13 What is the time when APNEA PRESSURE!! appears as a medium alarm
tone with associated flashing yellow LED? ___ sec (26-34)
( )
8.19.14 What is the time when the APNEA PRESSURE!!! appears as a high alarm
with flashing red LED and corresponding audible alarms? ___ sec (56-64)
NOTE: With the threshold low alarm enabled a corresponding threshold low alarm
shall appear.
8.19.15 After the APNEA PRESSURE!!! alarm is displayed as a high, slowly
increase the test pressure.
( )
8.19.16 At what pressure does the APNEA PRESSURE alarm deactivate? ___ cm
H
8.19.17 Adjust the threshold to 18 cm H
8.19.18 Increase the pressure to 20 cm H
stopwatch.
( )
8.19.19 What is the time when CONTINUOUS PRESSURE appears as a high
alarm? ___ sec (12-18)
( )
8.19.20 Decreasing the pressure slowly, what is the pressure at which the
CONTINUOUS PRESSURE alarm deactivates? ___ cm H
8.19.21 Increase the test pressure.
( )
8.19.22 At what pressure does the Airway Pressure High!!! alarm activate? ___ cm
H
8.19.23 Bleed the pressure.
8.19.24 Using a syringe in place of the squeeze bulb, slowly create a sub-
atmospheric pressure.
( )
8.19.25 At what pressure does the Pressure Negative!!! alarm activate? ___ cm
H
8.19.26 Return the pressure to zero.
8.19.27 Does the PRESSURE NEGATIVE alarm deactivate? ___(Y)
8.19.28 Using the syringe, increase the pressure to 20 cmH2O; then decrease the
pressure to zero.
8.19.29 Press the Autoset soft key.
8.19.30 Does the threshold limit adjust to within four (4) cmH2O of peak pressure?
___ (Y)
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
O (7-13)
2
O (38-42)
2
O (-2 to -8)
2
53
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)
O.
2
O, maintain the pressure, and start a
2
O (15-21)
2