Fabius Tiro
Fabius Tiro Service Log
Entries:
92
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Figure 6
Service Log Entries Screen
Scroll to the return arrow and press the rotary knob to return to the Service
Log category screen.
Tables of service log entries, error codes, and their descriptions can be found
in
Table 1
through
Table
7.
Table 1
Communication Errors
Code
Description
C001
Handler Busy
C002
Incoming Overflow
C003
Outgoing queue overflow Bad processor, external device incor-
C004
Message not registered
C005
Character not transmitted Bad processor, external device incor-
C006
Interrupt problem
C007
Continuous interrupts
C008
Excessive Breaks
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Possible Cause
Bad processor, external device incor-
rect operation
Bad processor, bad cable or connec-
tor, incorrect settings, external device
incorrect operation
rect operation
External communications error
rect operation
Bad processor
Bad processor
Bad processor, incorrect settings,
external device incorrect operation
Revision B Released