O2 Zero Calibration - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Diagnostics
6.3

O2 Zero Calibration

Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Figure 19 Pressure Calibration Screen
Press the Exit key to return to the Calibration category screen, or the Standby
key to return to the Standby screen.
When O2 is selected, the screen shown in
Follow the on-screen instructions to perform the calibration. The Store Zero
and Exit keys are located just below their legends at the bottom of the screen.
If the calibration process fails, a "value not stored" message is displayed in
place of the Stored zero cell A or cell B values.
6020.002
Fabius Tiro
Figure 20
appears.
Revision B Released
111