PMS PROCEDURE (continued)
8.22.7
Disconnect all test equipment.
( )

8.23 Oxygen Flush Valve

8.23.1
Press and release the O
8.23.2
Does the flow of oxygen stop immediately? __ (Y)
8.23.3
Connect a 12-inch hose to the inspiratory valve.
8.23.4
Occlude the bag mount with test plug.
8.23.5
Insert the sensor from a calibrated O
the inspiratory valve.
8.23.6
Close all flow control valves.
8.23.7
Press the O
( )
8.23.8
What is the O
8.23.9
Remove the O
8.23.10 Disconnect the fresh gas hose from the breathing system. Using the fresh
gas test connector (4117361), test terminal (4104389), and appropriate
adapter (4115087), connect to fresh gas hose and attach test terminal to in
port of test volumeter.
8.23.11 Press and hold the O
4.
( )
8.23.12 What is the oxygen flush flow rate? ___L/min.
8.23.13 Remove the test minute volumeter and test fixture, and reconnect the
fresh gas hose.
8.23.14 Turn the System Power switch to ON.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
FLUSH button.
2
FLUSH button.
2
concentration after the value stabilizes?___%O
2
sensor from the inspiratory dome and install the plug.
2
FLUSH button for 15 seconds; multiply the value by
2
Minimum: 35 L/min. @ 50 psi (domestic)
Minimum: 25 L/min. @ 40 psi (2.8 bar)
Maximum: 75 L/min. @ 87 psi (6.0 bar)
60