Adjustment and Calibration Procedures
4. Connect test pressure gauge (P/N 4114807) between machine's oxygen
pipeline inlet connector and the oxygen pipeline supply hose.
5. Open the O2 cylinder valve and set the System Power switch to ON.
6. Set the oxygen flow to 200 mL per min.
7. Close the oxygen cylinder valve.
8. As the pressure drops, the O2 SUPPLY alarm should activate when the
pressure is between 16 and 24 psi as shown on the test gauge.
9. If the alarm activates when the pressure is below 16 psi or above 24 psi,
turn the adjustment set screw (see
as necessary to bring the set point into the correct range.
NOTE
Turn the screw counter-clockwise to decrease the alarm activation.
10. Set the System Power switch to OFF.
11. Disconnect the test gauge.
12. Connect the pipeline hoses.
13. Perform the PMS Procedure.
6002.002
RETURN TO CD-ROM TABLE OF CONTENTS
Figure
3). Repeat the test and adjust
Revision B Released