Diagnostics
9

Pump On/Off

Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
The pump on/off softkey provides the ability to toggle the pump ON or OFF as
part of the troubleshooting and diagnostics process.
1. Access the 'Real Time Values' screen. Refer to
Appearance with PCB
2. Press the 'PUMP ON/OFF' softkey to toggle the pump between the On
and Off state.
NOTE
The pump label located at the bottom left corner of the 'Real Time Values'
screen will be highlighted in reverse when the pump has been activated.
When first entering the screen, the initial state of the pump is On.
6020.002
Real Time Values Screen
4118079.
Revision B Released