Fresh Gas Flow Calibration - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Adjustment and Calibration Procedures
7
Fresh Gas Flow Cali-
bration
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Figure 7
Pressure Calibration Screen
1. Set the main power switch to ON.
2. Access the main service screen (ref.
3. Scroll to Calibration and select Fresh Gas Flow to bring up the flow cali-
bration screen. See
Figure
4. Close all flow control valves.
5. Verify no fresh gas flow is present.
6. Touch the "Store Zero" softkey.
7. At completion of calibration "Stored Zero" shall indicate Pass or Fail.
8. Press the Standby key to return to the standby screen.
6002.002
Main Service Screen
8.
Revision B Released
Fabius Tiro
Section).
217