Recommendations; How To Use This Manual; General Troubleshooting Guidelines - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

General
1

Recommendations

2
How To Use This
Manual
3
General Trouble-
shooting Guidelines
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Because of the sophisticated nature of Draeger Medical, Inc. anesthesia
equipment and its critical importance in the operating room setting, it is highly
recommended that only appropriately trained and experienced professionals
be permitted to service and maintain this equipment. Please contact
DrägerService® at (800) 543-5047 for service of this equipment in North
America. For service in Europe, call 49 (451) 882-4222. For service in other
countries, call (215) 721-5402.
Draeger Medical, Inc. recommends that the Fabius Tiro be serviced at six
month intervals. Periodic Manufacturer's Certification agreements are avail-
able for equipment manufactured by Draeger Medical, Inc. Please contact us
for further information concerning these agreements.
Draeger Medical, Inc. products/material in need of factory repair shall be sent
to:
For North America and other coun-
tries except Europe:
DrägerService®
3124 Commerce
Drive Telford, PA 18969
U.S.A.
(Include RMA Number)
The manual is divided into several sections. The DIAGNOSTICS section
describes self-test and service diagnostics for checking the system functions.
An understanding of the on-board service capabilities is necessary before
any attempt is made to troubleshoot the unit. The TROUBLESHOOTING sec-
tion lists error codes and provides troubleshooting guides to assist the Tech-
nical Service Representative in locating the source of a problem. The
REPLACEMENT PROCEDURES section contains instructions for removal
and replacement of assemblies that are considered field-replaceable. The
ADJUSTMENT AND CALIBRATION PROCEDURES section contains the
field procedures needed to restore original system specifications. The Peri-
odic Manufacturer's Service (PMS) PROCEDURE section outlines the steps
required to verify the electrical, mechanical, and pneumatic safety of the unit
and also, identifies components requiring periodic replacement. The SPARE
PARTS section is provided for use as a reference only to obtain part numbers
and descriptions for parts and assemblies for replacement purposes. For
items not shown in the Spare Parts section, contact DrägerService®.
Troubleshooting the Fabius Tiro should always begin by communicating with
those who observed or experienced a problem with the unit. This may elimi-
nate unnecessary troubleshooting steps. Once a general problem is identi-
fied, refer to the flow charts in the Troubleshooting Section to determine the
proper corrective action to be taken.
After any component is replaced, verify that the unit is operating properly by
running the appropriate diagnostic procedure. The PMS PROCEDURE must
also be performed after any component is replaced.
6020.002
Fabius Tiro
For service in Europe:
Dräger Medical AG & Co. KGaA
Moislinger Allee 53-55
Reparaturannahme
23542 Lübeck
Germany
Revision B Released
3