Adjustment and Calibration Procedures
Figure 9
PEEP Valve
9
Vacuum Adjustment
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Figure 10 PEEP Valve Calibration Screen
The vacuum adjustment is located on the pneumatic assembly and is acces-
sible by removing the back cover of the controller housing.
6002.002
Fabius Tiro