PMS PROCEDURE (continued)
( )
8.8

Scavenger

8.8.1
AGS Scavenger
AGS Scavenger Cleaning
8.8.1.1
8.8.1.2
8.8.1.3
8.8.1.4
8.8.1.5
8.8.1.6
AGS Scavenger Functional Test
8.8.1.7
8.8.1.8
8.8.1.9
8.8.1.10 Verify vacuum waste disposal system at manifold is active.
8.8.1.11 Reinstall canister and adjust scavenger flow control valve; verify
8.8.2
Passive Scavenger
Passive Scavenger Cleaning
8.8.2.1
8.8.2.2
8.8.2.3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
Remove all scavenger hoses one at a time and drain all
accumulated moisture. Inspect all scavenger hoses for deterioration
and replace any worn hoses.
Disconnect the hospital vacuum source from the scavenger.
Remove the reservoir canister from the scavenger by partial turn
counter-clockwise of the canister.
Remove and inspect the silencer; replace if needed.
Remove the flowmeter from the scavenger by turning the mounting
nut counter-clockwise. Inspect the tube and clean with compressed
air if needed.
Reassemble the scavenger assembly, and reactivate the vacuum
source.
Verify vacuum hose is connected to one of the two gas supply
connections.
Verify the additional gas surplus connection is plugged.
Verify all hose connections to the scavenger are complete at all
destinations.
float moves freely between the upper and lower limits.
Inspect all scavenger hoses for signs of wear and deterioration.
Replace any worn hoses.
Remove the anti-occlusion cage from the passive scavenger by
unscrewing it.
Inspect the filter.
24