Wall Mount Service Access Procedure (If Applicable) - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Replacement Procedure
2
Wall Mount Service
Access Procedure (If
Applicable)
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1. Remove the Vaporizer (Refer to the
2. Make the necessary provisions to all monitoring devices, breathing sys-
tem, and accessories on the Core Unit to ensure the unit can pivot up to
180º.
3. Remove the unit pivot knob and washer located under the wall mount
bracket, then using an allen wrench, remove the service access screw.
Refer to
Figure
2.
4. Remove the cylinder wrench from the holder, then carefully swivel the
core unit to gain access to the rear of the unit.
5. Perform necessary service.
6. Pivot core module back into its original position and reinstall the service
access screw, pivot knob, and washer.
7. Reinstall all monitoring devices, breathing system vaporizer or accesso-
ries to the core unit that were previously removed, then reinstall the cylin-
der wrench in its holder.
6020.002
Fabius Tiro
Vaporizers
Section).