Figure 23 Spontaneous Breathing (Expiration) - Typical for Breathing System P/N 4118378 or 4118379
7.3
Volume/Pressure
Mode Ventilation
32
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
The expiratory gas flows from the lung through the expiratory hose, the flow
sensor, the PEEP control valve, the expiratory valve, the breathing bag, and
through the absorber. At the same time, new fresh gas flows into the breath-
ing bag.
When the breathing bag is full, any excess gas mixture flows through the
non-return valve into the anesthetic gas scavenging system.
Volume Control Mode: General
A prerequisite for volume control is that the patient is supplied with a sufficient
amount of fresh gas.
The APL bypass valve opens in volume mode, allowing excess gas to be
vented to the scavenging system regardless of the MAN-SPONT valve set-
ting.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description
Revision B Released