Figure 12 Lower Diaphragm Removal
26. Install the replacement diaphragm down over the piston; fold the lower lip
of the diaphragm (now facing up) over the piston and reinstall the retain-
ing ring in the inner groove of the diaphragm inside the piston. See
12
and
Figure
14.
NOTE
The gap in the retaining ring (circlip) shall be aligned opposite the mark on
top of the piston as shown in
units where the mark on the piston does not exist.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement Procedure
Figure
14. This note does not apply to later
Revision B Released