7.2.1.2 Gas Selection

120
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Figure 27 O2 Flowmeter Position Screen (Virtual Flowtubes)
To return to the Flowmeter selection screen, scroll to the return arrow and
press the knob.
When Gas Selection is chosen, the screen will appear as shown in
Turn the rotary knob to choose Gas Selection, and press the knob to highlight
the choices. Turn the rotary knob to select either 2 or 3, then press the knob
to enter that selection.
The gas selection number 2 or 3 shall match the total gas options pneumati-
cally configured on the unit. Otherwise a 'ERR' will occur on the fresh gas dig-
ital display, where applicable.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Revision B Released
Diagnostics