Audio Silence - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

FABIUS TIRO
( )
8.21.10 What is the time when APNEA FLOW!!! appears as a high alarm? ___ sec
(26 - 34)
8.21.11 Disconnect the respiratory volume sensor cord from the Flow sensor.
( )
8.21.12 Does the "FLOW SENSOR FAIL!" message appear as an Advisory with a
single audible tone? ___ (Y)
8.21.13 Does the APNEA FLOW!!! alarm disappear? (Y)
8.21.14 Do the numerical values associated with volume disappear? ___ (Y)
8.21.15 Reconnect the respiratory volume sensor cord to the flow sensor, and
expiratory limb of breathing circuit to breathing system.
8.21.16 Press Standby, confirm, and recalibrate the flow sensor.
8.21.17 Fresh Gas Low
8.21.17.1 Select Volume Control mode and confirm.
8.21.17.2 Disconnect the hose attached to the breathing system's inspiratory
( )
8.21.17.3 After 30 seconds verify the APNEA PRESSURE!!!, APNEA
8.21.17.4 Reconnect the patient Y connector to the patient inspiratory port and
8.21.17.5 Verify the FRESH GAS LOW!! alarm disappears after two (2)
( )

8.22 Audio Silence

8.22.1
Press the
8.22.2
Does the LED on the Silence Alarms key light? Is the audio alarm silenced
for 120 sec.?
8.22.3
Press Standby key and confirm.
8.22.4
Cycle the system power and wait for the Standby screen to appear.
8.22.5
Press Volume Control and confirm.
8.22.6
Verify the 120-sec. delay starts at power-up.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
port.
FLOW!!!, MINUTE VOLUME LOW!!!, FRESH GAS LOW!!, and NO
FRESH GAS!!! alarm messages appear on the display with a
corresponding audible alarm.
activate the flush.
seconds following the next ventilator cycle. ___ (Y)
key.
59
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)