Diagnostics
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Code
Description
C009
Excessive Framing
C010
Excessive Parity
C011
Excessive Overrun
C012
Non-existent UART
Table 2
Monitor Errors
Code
Description
M001
O2 <10%
M002
O2 LOW INOP
M003
O2 DIFF HIGH
M004
BATTERY <10%
M005
BATTERY <20%
FG01
N2O CAL ERR
FG02
AIR CAL ERR
FG03
O2 CAL ERR
Table 3
Diagnostic Errors
Code
Description
F001
Diag Watchdog
F002
Diag System Ram
F003
Diag Program Memory
F004
Diag Video Test
F005
Diag Interrupts
F006
Diag A/D Converter
F007
Diag NM Ram
F008
Diag Serial Port
F009
Diag Clock
F010
Diag Speaker
F011
Diag Main Power
6020.002
Fabius Tiro
Possible Cause
Bad processor, incorrect settings,
external device incorrect operation
Bad processor, incorrect settings,
external device incorrect operation
Bad processor, incorrect settings,
external device incorrect operation
Bad processor
Possible Cause
O2 cell not working properly
O2 cell inoperative or disconnected
Difference between Cell A and Cell B
>25%
Battery level below 10% of capacity
while operating without mains power
Battery level below 20% of capacity
while operating without mains power
N2O sensor value out of range
AIR sensor value out of range
O2 sensor value out of range
Possible Cause
Bad hardware
System RAM fails on pwr up, bad
RAM chip, bad CPU, bad trace, etc.
Bad flash download, bad flash chip
Video memory failed, bad video chip
Bad hardware
Bad A/D hardware
Corrupt entry on pwr up
Failed pwr up loopback test, bad
UART chip
Bad hardware
Speaker not connected on pwr up
Mains pwr not detected on pwr up
Revision B Released