8.9.6.10 Reconnect fresh gas hose to breathing system; connect ventilator
8.9.7
Leakage Control Port
8.9.7.1
8.9.7.2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
Verify the APL valve is in the MAN position and set to 70 cmH2O.
Slowly increase O2 flow to maintain 40 cmH2O on the digital
pressure meter.
Verify the flow on the 0-250 cc test flowmeter indicates less than 10
mL/min. (0.01 L/min.).
Close the O2 flow control valve.
Remove the occlusion plugs and test equipment from the breathing
system.
hose to vent port and breathing system.
Disconnect the APL and PEEP control hoses from the interface
panel at rear of machine.
Connect the PEEP control hose to the input of the 0-250 cc test
flowmeter.
35