Software Update Procedure

Software Update Proce-
dure
1

Requirements

2
Boot Strap Down-
load Procedure
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
This section outlines the equipment needed, Service Procedures for specific
software versions, and the BootStrap Download Procedure.
NOTE
The screen illustrations contained in this document are for reference only
and therefore may or may not reflect the current equipment or software ver-
sion being installed or exact Fabius Product.
Equipment required:
Cable Asm, Vitalink, Part No. 4110328 (9-pin to 9-pin). Alternate cable
combination: 7901808 and 7901762
IBM® PC or IBM PC Compatible configured with:
Windows 98, Windows NT, Windows 2000, Windows NT 4.0, or Windows
XP
RS-232C Serial port connected to COM 1 or COM 2
Parallel port (LPT1 or LPT2)
CD-ROM and Floppy Drive
Standard sound card
2 PCMCIA slots (on laptops)
Modem (or external modem) V34+ 28800 Baud
The following Service Procedure(s) is available via Lotus Notes and provides
detailed instructions to install/upgrade specific software versions. Refer to the
following Service Procedure or contact DrägerService - Technical Support:
Phone 1-800-4-Drager, Phone: 215-721-5402, or e-mail to techsup-
port@draegermed.com.
SP00297 - Software Version 2.0N to 2.N Upgrade
This procedure uses Version 2.N of the SEND.IMG program. V2 is a new
Win32 program that adds the Fabius Tiro boot Strap load function to the orig-
inal SENDIMG.EXE program that is currently used in existing products.
When a Fabius Tiro does not have a valid bootable image installed, it can not
perform an Automatic Image Reload Operation. The same PC and files are
used (including SENDIMG.EXE V2).
1. Record the market setting on the unit for reconfiguration of the controller
PCB after download. Refer to the
tion to access the appropriate service screen.
2. Set the machine power switch to the OFF position.
3. Remove the back panel and change the JP2 (on Control PCB P/N
4116632) or JP3 (on Control PCB P/N 4118079) connector on the control
PCB from the Boot Flash position
(Figure
2). Refer to or for proper connector locations.
6020.002
Fabius Tiro
General
portion of the Diagnostics Sec-
(Figure
1) to the Boot Strap position
Revision B Released
229