PMS PROCEDURE (continued)

8.15 Flowmeters

( )
8.15.1
Oxygen Flowmeter
8.15.1.1 Open the O
8.15.1.2 Is it possible to adjust the flow of oxygen to 10 L/min.? ___ (Y)
8.15.1.3 Close the O
8.15.1.4 Connect the O
8.15.1.5 Is the correct flow control knob and label attached to the oxygen
8.15.1.6 Close the oxygen flow control valve.
( )
8.15.2
Nitrous Oxide Flowmeter (if applicable)
8.15.2.1 Set the oxygen flow to 4 L/min.
8.15.2.2 Open the nitrous oxide cylinder valve.
8.15.2.3 Is it possible to adjust the flow of nitrous oxide to 10 L/min.? ___ (Y)
8.15.2.4 Close the nitrous oxide cylinder valve and bleed the pressure.
8.15.2.5 Connect the N
8.15.2.6 Is the correct flow control knob attached to the N
8.15.2.7 Close the oxygen and nitrous oxide flow control valves.
( )
8.15.3
Air Flowmeter
8.15.3.1 Connect the Air pipeline supply (if applicable).
8.15.3.2 Is it possible to adjust the flow of air to 10 L/min.? ___ (Y)
8.15.3.3 Close the air flow control valve and disconnect the Air pipeline
8.15.3.4 Is the correct flow control knob attached to the air flow control
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
cylinder valve.
2
cylinder valve and bleed the pressure.
2
pipeline supply, and verify the operation of the
2
oxygen flowmeter.
flow control valve? ___ (Y)
O pipeline supply, and verify the proper operation of
2
the N
O flow.
2
valve? ___ (Y)
supply.
valve? ___ (Y)
48