FABIUS TIRO
8.20.8
Pressure Limiting Valve Test
8.20.8.1 Set ventilator to Volume Control mode and confirm.
8.20.8.2 Set: pmax = 40 mbar (cmH2O)
8.20.9
Remove ventilator hose from ventilator and connect a digital manometer to
the ventilator outlet port.
( )
8.20.10 Verify relief valve opens between 70 and 80 mbar (cmH2O) when
ventilator is in inspiratory phase.
SV88022
DEVICE 4113119 (MODIFIED)
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
Vt = 600 mL
Freq = 6 BPM
TI:TE = 1:1
TIP:TI = 10%
PEEP = 0 mbar (cmH2O)
FRESH GAS LEAK TEST
57
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)