8.14.2.4 Depress and release the test device push button.
( )
8.14.2.5 What is the pressure on the test gauge when the "O2 SUPPLY
( )
8.14.2.6 Does the "O2 SUPPLY LOW!!!" message appear on the display?
8.14.2.7 Bleed the remaining O
8.14.2.8 Remove the test gauge from the machine and reconnect the
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
LOW!!!" message and associated red indicator appear? ___ PSI (16
- 24)
___ (Y)
flow control valve.
pipeline supply hose.
PRESSURE TEST ASSEMBLY
4114807 WITH ADAPTERS
O2
PIPELINE
HOSE
47
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)
pressure from the system, then close the
2