Fabius Tiro
7.1.1
Model Type
116
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Figure 23 System Settings Screen
When the Model Type category is selected, the screen shown in
will appear.
Turn the rotary knob to scroll down to select the model type 'Tiro', then press
the knob to confirm.
NOTE
Machines from the factory are configured with the appropriate model type.
Other than performing a software download or PCB replacement, the
model type should not be changed. Changing the model type will disable
some ventilator options previously configured and will require reconfigura-
tion.
To reconfigure the unit, contact DrägerService - Technical Support: Phone
1-800-4-Drager, Phone: 215-721-5402, or e-mail to techsupport@drae-
germed.com and provide the following information to receive the necessary
release codes:
–
Machine Type (Fabius GS or Fabius Tiro)
–
Feature Description and Part Number
–
Machine Serial Number
–
Device ID Number
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Revision B Released
Diagnostics
Figure 24