Table 5
Service Log Events
Code
Description
S006
Service Log Clear
SS90
test message
I001
CPU Diag Recovery
Table 6
User Interface Events
Code
Description
U001
New Device ID Created
U002
Option Enabled
U003
Option Disabled
U007
Model Changed to FGS
U008
Model Changed to Tiro
U009
Monitoring Disabled
U010
Monitoring Enabled
U012
SOS Recovered
U013
FGS SOS Invalid
U014
Tiro SOS Invalid
Table 7
Ventilator Errors
Code
Description
V001
VENT PEEP NOT CAL
V002
ZERO POS INVALID
V037
EX VALVE LEAK
6020.002
Fabius Tiro
Possible Cause
Log cleared by service tech, SW
upgrade
Factory test - no failure
CPU recovered from bad initial-
ization
Possible Cause
Machine detected a new device
ID was needed because of
NVRAM corruption or previously
did not exist
A secure option was enabled
A secure option was disabled
Model-type change
Model-type change
Monitoring-mode change
Monitoring-mode change
A Secure Option Setting was
recovered after a partial NVRAM
corruption
The FGS Secure Options Setting
NVRAM file was found corrupt
and no Options were recovered
The Tiro Secure Options Setting
NVRAM file was found corrupt
and no Options were recovered
Possible Cause
Auto ventilation was attempted
without valid PEEP calibration
Crossing of light barrier at bottom
vent not detected
Greater than 15 mL of Tidal Vol-
ume during inspiration, and this
Vt is greater than 10% of Vt flow
during expiration
Revision B Released
95