Fio2 Measurement - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Function Description
ITEM
1
Selects volume controlled ventilation mode. Refer to Operator's Manual
2
Select pressure controlled ventilator mode. Refer to Operator's Manual
3
Reserved for future use
4
Reserved for future use
5
Controls table lamp: Off/On
6
Places ventilator in Man Spont mode. Refer to Operator's Manual
7
Soft Keys: active the corresponding function that appears on screen above the key
8
For setting alarm limits. Refer to Operator's Manual
9
Setup Key: activates sub-screens for monitoring functions. Refer to Operator's Manual
10
Home key: returns display to main screen shown before standby
11
Rotary control: moves the cursor on the screen; confirms selection when pressed
12
Alarm Status Indicators: Flashing Red: Warning; Flashing Yellow: Caution; Solid Yellow: Advi-
sory
13
Alarm Silence key: silences all active alarms for two minutes
14
Power ON indicator: lighted when machine is plugged into an active AC outlet
15
Returns unit to Standby mode
13

FiO2 Measurement

Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Descriptions of the numbered items in
Table 3
Fabius Tiro Control Panel Key for
FUNCTION
The O2 sensor measures the fraction of inspired O2 (FiO2) in the respiratory
gas.
The O2 sensor contains an alkaline electrolyte, a lead anode, two gold cath-
odes, and a Teflon membrane. The spatial separation of the two gold cath-
odes allows a voltage comparison to be made as explained below.
The O2 sensor is an electrochemical cell that generates a voltage from the
ion current.
6020.002
Fabius Tiro
Figure 35
are given in the
Figure 35
Revision B Released
Table
3.