FABIUS TIRO
8.17.1.3 Does the oxygen monitor read 97-100% after the value
8.17.1.4 Set the nitrous oxide flow to 2 L/min.
( )
8.17.1.5 After the value stabilizes on the oxygen sensor, what is the O
8.17.1.6 If measured value is within the range, close the N2O flow control
( )
8.17.1.7 If measured value is not within range, close the nitrous oxide flow
( )
8.17.1.8 Set nitrous oxide flow to 2.5 L/min and oxygen flow to 2.5 L/min on
8.17.1.9 Close the nitrous oxide flow control valve.
8.17.2
Oxygen + Air Concentration
8.17.2.1 Depress the O
8.17.2.2 Does the oxygen monitor read 97-100% after the value
8.17.2.3 Set the air flow to 2 L/min.
( )
8.17.2.4 After the value stabilizes, what is the O
8.17.2.5 Close the air and O2 flow control valves.

8.18 SORC

8.18.1
Depress the O
8.18.2
Set the O2 and N2O flow to 10 L/min.
8.18.3
Set the O2 flow control valve to 0.8 L/min.
( )
8.18.4
What is the oxygen concentration after the value stabilizes? ___% (
8.18.5
Adjust the oxygen flow to 1.5 L/min.
( )
8.18.6
What is the oxygen concentration after the value stabilizes? ___% (
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
stabilizes?__(Y)
concentration? ___% (62-72)
valve and proceed to Section 8.18.2.
control valve and adjust the oxygen flow until total flow meter reads
4 L/min. What is the corresponding display flow rate? ___ L/min (3.9
- 4.7)
the displays. What is the corresponding flow visually approximated
on the total flow meter? ___L/min (4.5 - 5.5)
FLUSH button for 5 seconds.
2
stabilizes?__(Y)
77)
FLUSH for 5 seconds.
2
51
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)
concentration? ___%(71-
2
2
23%)
23%)