PMS PROCEDURE (continued)
8.19.31 Press the Man Spont key and confirm.
8.19.32 Reconnect the breathing pressure line to the breathing system.

8.20 Ventilator

( )
8.20.1
Manual Ventilation
8.20.1.1 Cycle machine power-using On/Off switch and allow completion of
8.20.1.2 Verify the ventilator switches to standby mode after completion of
8.20.1.3 Connect a test lung (8401892) to the breathing circuit Y-piece, and
8.20.1.4 Attach a breathing bag to the breathing bag connector of the
8.20.1.5 Select Man Spont.
8.20.1.6 Set the fresh gas flow to 3 L/min.
8.20.1.7 Set the APL valve to MAN, 30 cmH2O.
8.20.1.8 Verify that manual breathing is possible by manually squeezing the
( )
8.20.2
Spontaneous Breathing
8.20.2.1 Set the APL valve to SPONT.
8.20.2.2 Verify that spontaneous breathing is possible with the test lung. ___
8.20.2.3 Close O2 flow control valve.
( )
8.20.3
Flow Sensor Zeroing
8.20.3.1 Select Standby mode and confirm.
8.20.3.2 Disconnect the expiratory hose from the breathing system.
8.20.3.3 Press the Calibrate Flow Sensor key and perform the on-screen
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
the self-test.
self-test.
attach to breathing system.
breathing system.
breathing bag. ___ (Y)
(Y)
instructions. At completion of calibration verify "Flow Calibration
Completed" message appears.