FABIUS TIRO
8.20.3.4 Reconnect the breathing circuit to the inspiratory and expiratory
( )
8.20.4
Ventilator Delivery
8.20.4.1 Set APL to Man position and set to 30 cmH2O.
8.20.4.2 Switch to Volume Control mode and set the fresh gas flow to 3 L/
8.20.4.3 Interconnect a test volumeter (P/N 2212300 w/4115087 connectors)
8.20.4.4 Is Volume Mode displayed and flashing? ___ (Y)
8.20.4.5 Depress Flush momentarily to inflate bag.
8.20.4.6 Set: pmax = 70 mbar (cmH2O)
8.20.4.7 Press the rotary knob.
8.20.4.8 Is Volume Control displayed? ___ (Y)
8.20.4.9 Verify that ventilation starts. ___ (Y)
8.20.4.10 Does drive run quietly and smoothly ___ (Y)?
( )
8.20.4.11 Verify display Vte = 300 - 450 mL (___ mL/min.)
( )
8.20.4.12 Verify Minute Volume on test volumeter and MV on display are
8.20.5
PEEP Pressure Accuracy
8.20.5.1 After 10 breaths, press PEEP parameter button.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
ports of the breathing system.
min.
between the expiratory limb of breathing system and breathing
circuit.
Vt = 380 mL
Freq = 12 BPM
TI:TE = 1:1
TIP:TI = 10%
PEEP = 0 cmH2O
within 20% of each other.
55
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)