Dräger Medical Fabius Tiro Technical & Service Manual page 281

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

FABIUS TIRO
8.11.1.4 Reconnect the O2 pipeline hose to the O2 pipeline inlet connector
8.11.1.5 Set the O2 flow to 4 L/min.
( )
8.11.1.6 After the digital manometer stabilizes, what is the regulator output
8.11.1.7 Deplete O2 pressure from the pipeline supply.
8.11.1.8 Close the flow control valve and disconnect the O2 pipeline hose
8.11.1.9 Remove test equipment and reconnect the O2 pneumatic hose
8.11.2
N2O Inlet Regulator
8.11.2.1 Disconnect the N2O pneumatic hose from the outlet connector on
8.11.2.2 Interconnect a digital manometer between the N2O outlet connector
8.11.2.3 Reconnect the O2 and N2O pipeline hoses to the inlet block and
( )
8.11.2.4 Set the O2 and N2O flows to 4 L/min. After the digital manometer
8.11.2.5 Deplete the O2 and N2O pipeline pressures.
8.11.2.6 Close the O2 and N2O flow control valves and disconnect both
8.11.2.7 Remove test equiupment and reconnect the N2O hose from the flow
8.11.3
Pipeline Check Valves
O2 Pipeline Check Valve:
8.11.3.1 Attach a Swagelock fitting (P/N 4109318) to the inlet of the S000081
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
and pressurize the O2 supply.
pressure? ___psi (28 - 32 psi)
from the inlet.
leading from the O2 flow control valve to the O2 connector on the
inlet block.
the gas inlet block that connects to the N2O flow control valve.
and the hose removed in the previous step. See illustration on
previous page.
activate the pipeline supplies.
stabilizes, what is the N2O regulator output pressure? ___psi (28 -
32 psi)
pipeline hoses from the inlet block.
control valve to the N2O outlet on the inlet block.
test flowmeter.
39
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)