Dräger Medical Fabius Tiro Technical & Service Manual page 62

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Electromagnetic Testing and Compliance
56
Dräger Medical Systems, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Immunity
IEC 60601
test
test level
Conducted
3 Vrms
RF
IEC61000-4-
150 kHz to
6
80 MHz out-
side ISM
bands (ref.
NOTE A)
10 Vrms
150 kHz to
80 MHz in
ISM bands
(Ref. NOTE
A)
Radiated RF
10 V/m
IEC61000-4-
80 MHz to
3
2.5 GHz
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Compliance
Electromagnetic environ-
level
ment-guidance
Recommended separation
distance
[V1] V
d=[3.5/V1]vP
Complies
[V2] V
d=[12/V2]vP
Complies
[E1] V/m
d=[12/E1]vP 80 MHz to
Complies
800 MHz
d=[23/E1]vP 800 MHz to
2.5 GHz
where P is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufac-
turer and d is the recom-
mended separation distance
in meters (m) (Ref. NOTE B).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey (ref. NOTE C), should
be less than the compliance
level in each frequency range
(Ref. NOTE D).
Interference may occur in the
vicinity of equipment marked
with the following symbol:
((e»)
Revision B Released
Fabius Tiro