FABIUS TIRO
8.7

Site Configurations

8.7.1
Scroll to Configure and confirm.
O2 Position (virtual flowtubes):
( )
8.7.2
Scroll to Flowmeter and confirm. Scroll to O2 Position and confirm. Select
position of virtual flowmeter (left or right) in accordance with local
requirements or customer demand. Press to confirm. record position on
PMS form.
Gas Selection:
( )
8.7.3
Press to confirm, scroll to Gas Selection and confirm. Verify Gas Selection
type (2 or 3) is in accordance with machine configuration. Press to confirm
(x2). Record selection type on PMS form.
Flowtube Resolution:
8.7.4
Scroll to Flowtube Res. and press to confirm. Scroll to change state and
confirm. Scroll to change state in accordance with customer demand.
O2 Whistle:
( )
8.7.5
Press to confirm, scroll to O2 whistle. Verify O2 whistle selection (enabled
or disabled) is in accordance with local requirements or customer demand.
Press to confirm (x2). Record position on PMS form.
Alarms:
( )
8.7.6
Scroll to No Fresh Gas and confirm. Verify state of No Fresh Gas alarm is
enabled. Press to confirm (x2). Record position on PMS form.
( )
8.7.7
Scroll to Fresh Gas Low Alarm. Verify Fresh Gas Low Alarm position
(enabled or disabled) is set to enabled position. Press to confirm (x2).
Record position on PMS form.
( )
8.7.8
Scroll to Threshold Low and confirm. Verify state of Threshold Low Alarm
is enabled. Press to confirm (x2). Record position on PMS form.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
APPENDIX A
21
PMS PROCEDURE (continued)