Dräger Medical Fabius Tiro Technical & Service Manual page 270

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

PMS PROCEDURE (continued)
8.9.2.12 Push the System Power switch ON ( ).
8.9.2.13 Open the O
8.9.2.14 Push the System Power switch OFF (
8.9.2.15 Reconnect the fresh gas connector from the machine to the
( )
8.9.3
Breathing System Leak
8.9.3.1
8.9.3.2
8.9.3.3
8.9.3.4
8.9.3.5
8.9.4
APL Valve Verification
8.9.4.1
8.9.4.2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
flow control valve to 5 L/min., purge the system for 5
2
seconds, then close the O
freshgas outlet connector on the breathing system.
Verify the System Power switch is On.
Press the Leak/Compl Test softkey.
Close all flow control valves.
Perform the on-screen instructions.
Verify all applicable tests indicate a Pass condition.
Connect a digital manometer between the inspiratory and expiratory
ports of the breathing system (using two test terminals and modified
4113119 fresh gas leak test device). See following illustration.
Attach a breathing bag to the bag terminal connector on the
breathing system and install a test cap on the ventilator port.
28
RETURN TO CD-ROM TABLE OF CONTENTS
flow control valve.
2
).