O2 Whistle Language Date Format Time Format O2 Position
USA
Disabled
178
Dräger Medical AG & Co. KGaA
NOTE:
CONTROL PCB P/N 4118079
Verify jumper at JP2 is installed on both pins. This activates the 3V lithium
battery (BT1) for the non-volitile RAM.
8. Reconnect all of the cables and hoses that were previously disconnected.
NOTE
Replacement PCB's come with a new hose attached to the LS2 pressure
sensor. Remove and discard the original 2-hose assembly from the breath-
ing pressure interface panel connector inside the unit and attach the new
hose to the breathing pressure interface connector.
9. Reinstall the back panel on the controller housing.
10. Restore power to the machine and observe the power-up diagnostic dis-
play (see the
Diagnostics
working properly.
11. Access the appropriate service screen and update the machine serial
number, clear service log and reset hours run. The machine serial num-
ber is located on the rear of the machine.
12. Perform the market settings per the original machine configuration or cus-
tomer requirements.
NOTE
The default "Market Selection" for the replacement PCB is "U.S." To
change the language and pressure measurement units, the "Market Selec-
tion" must be changed to "Non-U.S." (see the Diagnostics Section). Refer
to
Table 1
for the market settings that correspond to your region.
13. If Applicable, perform
Adjustment and Calibration Section.
14. If applicable, enter any previously recorded release codes to restore any
ventilator feature option(s), or contact DrägerService - Technical Support:
Phone 1-800-4-Drager, Phone: 215-721-5402, or e-mail to techsup-
port@draegermed.com and provide the following information to receive
the necessary release codes:
Machine Type (Fabius GS or Fabius Tiro)
Feature Description and Part Number
Machine Serial Number
Device ID Number
15. Perform the PMS procedure, including all calibrations and site configura-
tions.
Table 1
Market Settings
English
M/D/Y
6020.002
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement Procedure
Section) to verify that the replacement PCB is
Pump Calibration
procedure in accordance with
12 Hour
Right
Revision B Released
Alarm
Pres.
Sounds
Units
Drager
cmH2O