Dräger Medical NeoFlow Addendum To Operating Instructions

Dräger Medical NeoFlow Addendum To Operating Instructions

Neonatal patient mode for evitaxl

Advertisement

Quick Links

NeoFlow
Neonatal Patient Mode
Addendum to
Operating Instructions
Evita
*
XL
*
These Operating Instructions apply also
to Evita 4 with the XL option
D
M E D I C A L

Advertisement

Table of Contents
loading

Summary of Contents for Dräger Medical NeoFlow

  • Page 1 M E D I C A L ™ NeoFlow Neonatal Patient Mode Addendum to Operating Instructions Evita These Operating Instructions apply also to Evita 4 with the XL option...
  • Page 2 T T T T r r r r a a a a d d d d e e e e m m m m a a a a r r r r k k k k s s s s The Dräger name and logo are registered trademarks of Dräger. AutoFlow® is a registered trademark of Dräger. NeoFlow™ is a trademark of Dräger. © Dräger Medical AG & Co. KGaA, 2002 All rights reserved, Subject to modifications...
  • Page 3: Table Of Contents

    Contents Contents Important Safety Information ......4 Care ..........32 Operator's Responsibility for Patient Safety .
  • Page 4: Important Safety Information

    Important Safety Information Operator's Responsibility for Patient Safety Limitation of Liability Important Safety Information Limitation of Liability Operator's Responsibility for Patient Safety Draeger Medical, Inc.'s liability, whether arising out of or WARNING ! related to manufacture and sale of the goods, their installation, demonstration, sales representation, use, performance, or Strictly follow this Operator's Instruction Manual otherwise, including any liability based upon Draeger Medical,...
  • Page 5: Warranty

    Important Safety Information Warranty Warranty All Draeger products are guaranteed to be free of defects for a Application of this warranty is subject to the following period of one year from date of delivery. conditions: The following are exceptions to this warranty: 1.
  • Page 6: Definitions

    Important Safety Information Definitions Definitions WARNING ! A WARNING statement refers to conditions with a possibility of personal injury if disregarded. CAUTION ! A CAUTION statement designates the possibility of damage to equipment if disregarded. N N N N O O O O T T T T E E E E : : : : A NOTE provides additional information intended to avoid inconveniences during operation.
  • Page 7: Summary Of Warnings And Cautions

    WARNING ! flammable gases or anesthetics. This device is neither approved nor certified for use in Installation of the NeoFlow option into Evita ventilators areas where combustible or explosive gas mixtures are may only be performed by factory trained and authorized likely.
  • Page 8 Important Safety Information Summary of WARNINGS and CAUTIONS WARNING! CAUTION ! Effect of aerosols on sensors, filters, Do not ventilate larger pediatric patients with serious and heat and moisture exchangers! infections and a severe cough using the neonatal flow The measuring function of the flow sensor may be sensor.
  • Page 9 Important Safety Information Summary of WARNINGS and CAUTIONS Precautions during maintenance WARNING ! To avoid any risk of infection, clean and disinfect ventila- tor and accessories before any maintenance according to established hospital procedures - this applies also when returning ventilators or parts for repair. WARNING ! Preventive Maintenance work on the Evita ventilators...
  • Page 10: Intended Use

    Intended Use Restrictions of Use Intended Use NeoFlow – neonatal mode with base flow. Extends the patient range of the Evita intensive care ventilator to infants and premature babies with a minimum body weight of 1.1 lbs (0.5 kg). Extends the range of flow monitoring with Evita during pediatric and neonatal ventilation employing a proximal flow...
  • Page 11: Preparation

    Before Using for the First Time Preparation Before Using for the First Time Installing the NeoFlow option WARNING ! Installation of the NeoFlow option into Evita ventilators may only be performed by factory trained and authorized service personnel. Follow installation instructions in the respective documentation.
  • Page 12: Preparing For Use

    Preparation Preparing for Use Preparing for Use Installing the neonatal flow sensor Prepare patient circuit – see chapter "Ventilating Infants" in the Evita Operating Instructions. Only the neonatal flow sensor (84 11 130) may be used. WARNING ! Do not use the Y-piece with integral flow sensor (part no.
  • Page 13: Device Check

    Preparation Device Check Device Check The NeoFlow option expands the Evita pre-use check procedure by the following function: — Calibration neo. flow sensor...
  • Page 14: Calibrating The Neonatal Flow Sensor

    Starting calibration for EvitaXL Press » Sensor Parameter« key. The »Sensor Parameter«, menu is displayed. Select »NeoFlow« menu, flow monitoring is activated. Start calibration: Touch »Start« screen key. The screen key turns green, the ventilator now calibrates the flow sensor.
  • Page 15: Replacing The Neonatal Flow Sensor

    Preparation Replacing the Neonatal Flow Sensor Calibration procedure Remove ET-tube connector, Remove neonatal flow sensor from Y-piece, Insert patient ET-tube connector directly into the Y-piece for the duration of the calibration. WARNING ! The operator of the ventilator must still assume full responsibility for proper ventilation while flow monitoring is not available during calibration.
  • Page 16: Operation

    Operation Selecting Neonatal Mode Operation Selecting Neonatal Mode The required patient mode can be selected from the Evita menu immediately after switching on or while in standby mode: Adult« » = Adult » Ped.« = Pediatric » Neo.« = Neonatal The menu ranges can be configured, see "Configuring Ventilation, Setting the patient range", on page 28.
  • Page 17: Volume Controlled Ventilation In Neonatal Mode

    Operation Volume Controlled Ventilation in Neonatal Mode Volume Controlled Ventilation in Neonatal Mode ® The AutoFlow ventilation mode supplement is always active when using volume controlled ventilation (CMV, SIMV, MMV) in neonatal mode. ® AutoFlow – for automatic regulation of "Insp.
  • Page 18: Back-Up Ventilation In Neonatal Mode

    Operation Back-up Ventilation in Neonatal Mode Pressure Support Ventilation (PSV) Back-up Ventilation in Neonatal Mode during volume-controlled neonatal ventilation Volume-controlled ventilation during neonatal ventilation is only possible with intact flow monitoring. If neonatal flow monitoring fails or has been switched off during volume- controlled ventilation, Evita automatically switches over to pressure-controlled back-up ventilation.
  • Page 19: Apnea Ventilation In Neonatal Mode

    Operation Apnea Ventilation in Neonatal Mode Apnea Ventilation in Neonatal Mode Unlike in adult or pediatric ventilation, pressure controlled apnea ventilation is started after the set alarm time > ) if an apnea occurs in Apnea ventilation modes with activated apnea ventilation.
  • Page 20: Niv (Mask Ventilation)

    Operation NIV (Mask Ventilation) NIV (Mask Ventilation) See Evita Operating Instructions for details on Non Invasive Ventilation (NIV). Set ventilation parameters As described for »Tube« application mode. An additional screen knob »T « is insp displayed for CPAP/P Supp In patient modes »Pediatric« or »Neonate«, Evita limits the maximum duration of a pressure support breath to...
  • Page 21: Flow Monitoring During Neonatal Ventilation

    Evita displays the »Sensor Parameter« menu. Touch screen key for the sensor to be deactivated, in this case »NeoFlow«. Touch »Off« screen key, the screen key turns yellow. Press dial knob to confirm, the screen key turns green. Neonatal flow monitoring is deactivated, and the corresponding measured values disappear from the screen.
  • Page 22: Flow Monitoring During Pediatric Ventilation

    Operation Flow Monitoring During Pediatric Ventilation Flow Monitoring During Pediatric Ventilation If the neonatal flow sensor is present and intact during ventilation in pediatric patient mode, it will perform the function of pediatric flow monitoring if neonatal flow monitoring is activated. If the neonatal flow sensor is defective or if neonatal flow monitoring is deactivated, flow monitoring is performed by the expiratory flow sensor installed in Evita...
  • Page 23: Nebulizing Pharmaceutical Aerosols

    Operation Nebulizing Pharmaceutical Aerosols Nebulizing Pharmaceutical Aerosols In neonatal mode, it is only possible to nebulize pharmaceutical aerosols during pressure controlled ventilation. WARNING! Effect of aerosols on sensors, filters, and heat and moisture exchangers! The measuring function of the flow sensor may be impaired.
  • Page 24 Operation Nebulizing Pharmaceutical Aerosols Removing the neonatal flow sensor WARNING! The wires of the flow sensor are hot. If the neonatal flow sensor is left in the patient circuit for some time during nebulizing without being cleaned, deposits from the medicated aerosols may build up and impair flow measu- rement.
  • Page 25 Operation Nebulizing Pharmaceutical Aerosols 1 Remove corrugated silicone tubing of the patient circuit from the inspiratory adapter of the Y-piece and re-connect it to the input adapter of the nebulizer. 2 Using the Y-piece inspiratory adapter, connect the free end of the corrugated tubing to the nebulizer.
  • Page 26 Operation Nebulizing Pharmaceutical Aerosols Starting nebulization Deactivate neonatal flow monitoring, page 21. Touch »Special Procedure...« screen function key, Evita opens the »Additional Function« menu. Touch » Nebulizer« screen key, the key turns yellow. Press dial knob to confirm, the screen key turns green. The nebulizer is in operation.
  • Page 27: O 2 Concentration When Using Nebulizer

    Operation O2 Concentration When Using Nebulizer Oxygenation for Bronchial Suction Concentration When Using Nebulizer The nebulizer for medication aerosols operates continuously, when activated in the neonatal range. During nebulizer treatment, the base flow is increased from 6 L/min to 9 L/min. Depending on the set O concentration, the nebulizer is supplied with medical compressed air, oxygen, or a mixture of...
  • Page 28: Configuration Of Ventilation

    Configuration of Ventilation Setting the Patient Range Configuration of Ventilation Setting the Patient Range Select the required patient mode from the list on the configuration page, see Evita Operating Instructions. The following ranges are available: Adults only Pediatrics only Neonates only Adults or Pediatrics Pediatrics or Neonates Adults, Pediatrics, or Neonates...
  • Page 29: Start-Up Defaults For Ventilation Parameters And

    Configuration of Ventilation Start-up Defaults for Ventilation Parameters and Alarm Limits Start-up Defaults for Ventilation Parameters and Alarm Limits Start-up default values, i.e. the values for tidal volume V and ventilator rate f activated whenever Evita is switched on, may be defined as a function of either ideal body weight or of the patient mode.
  • Page 30 Configuration of Ventilation Start-up Defaults for Ventilation Parameters and Alarm Limits To select the initial values for V and f as a function of the ideal weight, the Radford nomogram has been extended to an ideal weight of 0.5 kg: Factory-set default Hospital-set default Weight...
  • Page 31: Defining Start-Up Defaults For Tube Compensation (Atc)

    Configuration of Ventilation Defining Start-up Defaults for Tube Compensation (ATC) Defining Start-up Defaults for Tube Compensation (ATC) Press » System Setup« function key. Touch »Therapy« screen key. Touch »Mode & Settings« screen key and enter access code 3032. Touch »Add. settings...« screen key. Touch the respective screen key to switch ATC on or off.
  • Page 32: Care

    Care Dismantling the Neonatal Flow Sensor Disinfecting/Cleaning/Sterilizing Care Dismantling the Neonatal Flow Sensor 1 Unplug flow sensor cable from both the sensor and the back panel of Evita 2 Remove the sensor element: Press the buttons on both sides while pulling the sensor out of the housing at the same time.
  • Page 33 Care Disinfecting/Cleaning/Sterilizing Flow sensor cable Disinfect by wiping with disinfectant based on the recommended active ingredients. Follow instruction labels with respect to prescribed concentrations and the necessary exposure times. CAUTION ! Do not allow any liquid into the connector of the flow sensor cable.
  • Page 34: Maintenance

    Evita ventilators. Maintenance Intervals Preventive maintenance Every 6 months by trained and factory authorized service personnel. The Evita NeoFlow option is serviced as part of the scheduled preventive maintenance of the Evita ventilator every six months.
  • Page 35: Troubleshooting

    In the table below, the additional messages specific to order of priority. NeoFlow are listed in alphabetical order. The table can be used as reference for rapidly identifying and remedying the For example, if two faults are detected at the same time, the cause of any alarm.
  • Page 36: Technical Data

    Technical Data Neonatal Settings Technical Data For neonatal mode, supplements Technical Data section in the Evita Operating Instructions. Neonatal Settings Tidal volume V Range 3 to 100 mL, BTPS* Resolution 1 mL Accuracy greater of ±8 % of set value or 1 mL Trigger sensitivity Range 0.3 to 5 L/min...
  • Page 37: Performance Data

    Technical Data Performance Data Display of Measured Values Performance Data Control principle Base flow with demand system, pressure-controlled, time-controlled Base flow 6 L/min (this can be changed by DraegerService to 9 L/min) Base flow during medicament nebulisation 9 L/min Insp. Flow up to 30 L/min Exp.
  • Page 38: Monitoring

    Technical Data Monitoring Materials Used Tidal volume V Range 0 to 999 mL, BTPS 1000 to 4000 mL, BTPS Resolution 0.1 mL 10 mL Accuracy greater of ±8 % of the measured value or 1 mL Tidal volume V Range 0 to 999 mL, BTPS 1000 to 4000 mL, BTPS Resolution...
  • Page 39: Theory Of Operation

    Theory of Operation – Special Features of Neonatal Ventilation Measuring Leakage Flow Theory of Operation – Special Features of Neonatal Ventilation Measuring Leakage Flow A small amount of respiratory gas almost always escapes between the tracheal wall and endotracheal tubes when ventilating newborns and infants with uncuffed tubes.
  • Page 40 Theory of Operation – Special Features of Neonatal Ventilation Measuring Leakage Flow Patient flow and tidal volume are then calculated as follows: Inspiration: Flow = Flow – Flow Patient, insp insp Leak ∫ Flow Patient, insp Expiration: Inspiratory flow Flow = Flow + Flow Patient, exp...
  • Page 41: Airway Pressure Measurement

    Theory of Operation – Special Features of Neonatal Ventilation Airway Pressure Measurement Airway Pressure Measurement Evita measures airway pressure indirectly by means of two internal pressure sensors in the apparatus. The sensors are installed in the inspiratory and expiratory lines, thereby eliminating the need for an external pressure measuring line between the Y-piece and the ventilator.
  • Page 42: Trigger Response

    Theory of Operation – Special Features of Neonatal Ventilation Trigger Response Trigger Response In neonatal mode, Evita detects a patient's spontaneous breathing using the neonatal flow sensor proximal to the patient. When spontaneous inspiration is detected, a synchronized, mechanical, and pressure controlled breath, or a pressure support breath is triggered according to the selected mode of ventilation.
  • Page 43: Autoflow

    Theory of Operation – Special Features of Neonatal Ventilation ® AutoFlow ® AutoFlow AutoFlow is a ventilation mode extension that optimizes flow control during mandatory breaths in the volume constant > ventilation modes CMV, SIMV, and MMV. In neonatal mode, AutoFlow is always active in all volume = f (V insp controlled ventilation modes (CMV, SIMV, MMV).
  • Page 44 Theory of Operation – Special Features of Neonatal Ventilation ® AutoFlow > Tidal volume V can be limited by the alarm limit »V V V V «. If the set alarm limit is exceeded once, Evita will generate an > advisory message (!).
  • Page 45: Glossary

    Glossary Abbreviations Glossary Abbreviations Abbreviation Explanation Mechanical ventilation frequency (setting) Spontaneous breathing rate (measured value) Flow Displayed real-time waveform, patient flow (measured value), with correction for leak Flow Flow through the expiratory valve during inspiration Flow Base flow (system setting), see technical data page 36 Flow Inspiratory flow, without correction for leak...
  • Page 46: Ordering Information

    Ordering Information Abbreviations Ordering Information Name/Description Article No. NeoFlow kit 84 13 563 consists of Extension PCB "Pediatric Flow" Neonatal flow sensor cable Reusable neonatal flow sensor housing ISO 15 "Water trap" kit for expiratory valve (2x) Pediatric cuvette for CO...
  • Page 47: Index

    Index Index Operation ........16 Abbreviations .
  • Page 48 These Instructions for Use apply only to Evita with Serial No.: If no Serial No. has been filled in by Dräger, these Instructions for Use are provided for general information only and are not intended for use with any specific machine or device.

This manual is also suitable for:

84 13 563

Table of Contents