Dräger Medical Fabius Tiro Technical & Service Manual page 104

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Fabius Tiro
98
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Figure 8
Real Time Values Screen Appearance with PCB 4118079
Press the knob to return to the Service Log category screen.
Table 8
Real Time Values Screen Key and Acceptable Range Data
Label
Description
Paw
(AD) Pressure Sensor Calib.
value
F1
(AD) Flow Sensor Calib.
value
O2
O2 Sensor Channel (in mV)
Channel 1 + 2
CAL
Calibration State of O2 Sen-
sor and Calibration value
h
Operating Hours (full hours) N/A
U+24
Supply voltage in V
Pvent
Ventilator Pressure
PP
Pump Vacuum Pressure
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Acceptable Range
Max 825 (ambient) Min 650
(ambient)
0 - 4095
OFFSET Min 39 - 9,64 mV
(21%) / OFFSET Max 541
132,00 mV(100%)
Average stored valve
between OFFSET Min and
OFFSET Max
Max - 29 Volts Min - 21 Volts
Max 825 (ambient) Min 650
(ambient)
0- 120 mB (Pump OFF) 120
- 270 mB (Pump ON)
Revision B Released
Diagnostics